Printer friendly page

iQ®200 Series Automated Urine Microscopy Analyzer Technical Specifications

Particles identified
by Auto-Particle
Recognition (APR™)
RBC
WBC
WBC clumps
Squamous epithelial cells
Non-squamous epithelial cells
Hyaline casts
Unclassified Casts
Crystals
Yeast
Bacteria
Sperm
Mucus
26 Particle
sub-classifications
Unclassified casts
— Granular
— Cellular
— Waxy
— Broad
— RBC
— WBC
— Epithelial cell
— Fatty
Crystals
— Calcium oxalate
— Triple phosphate
— Calcium phosphate
— Uric acid
— Calcium carbonate
— Leucine
— Cystine
— Tyrosine
— Amorphous
Non-squamous Epithelial Cells
— Renal epithelial
— Transitional epithelial
Yeast
— Budding yeast
— Yeast with pseudohyphae
Other
— RBC Clumps
— Fat
— Oval fat bodies
— Trichomonas
Specimen Processing Barcode or keyboard entry of ID;
60-specimen walk away capability;
10-tube rack system with continuous feed
Specimen Volume Min. volume: 2.0 mL of un-spun urine
Aspiration volume: approx 1.0 mL
Measurement Principle Flow Cell Digital Imaging
Particle Classification Auto-Particle Recognition (APR™) using advanced neural network software
Sample Throughput 40 / 70 / 101 samples per hour
Workstation Pentium Processor with flat panel monitor / keyboard / mouse
Sample Rack 10-tube rack
Data Storage Onboard storage of up to 10,000 patient results
Interface Bi-Directional with host query
Electrical Power Req. Microscopy Module: 90-240 VAC 50-60 Hz
Monitor: 100-240 VAC
Dimensions Microscopy Module: 22H x 21W x 24D in. (55.9H x 53W x 61D cm)
CPU: 13.5H x 5.5W x 15D in. (34.3H x 14W x 38.1D cm)
Monitor: 17.52H x 17.52 W x 8D in. (44.5H x 44.5W x 20.3D cm)
Printer: 10.1H x 13.8W x 17.7D in. (25.654H x 35.052W x 45D cm)
Load / Unload Stations: 9H x 13.5W x 19D in. (22.9H x 34.3W x 48.3D cm)
Weight Microscopy Module: 100 lbs. (46 kg)
CPU: 28 lbs. (13 kg)
Monitor: 19.1 lbs. (8.7 kg)
Load / Unload Stations: 12lbs. (5.5kg)

 

 

iChem®VELOCITY™ Urine Chemistry Analyzer Technical Specifications

Tests Performed 13 (10 Urine Chemistries, Specific Gravity, Color and Clarity)
Measurement Methods Test Strip: Three wavelengths - Iris Chromaticity with variable endpoint
Specific Gravity: Refractive index
Color / Clarity: In-flow transmission and scattering color sensor with white light illumination
Measurement Wavelengths 472, 520, 630 nm
Measurement Principle Reflectance Photometry LEDs with CMOS Camera
Test Strip iChem Urine Chemistry Strip, 100-test vial; Storage 34° - 86°F (1° - 30°C)
Test Strip Capacity 1 to 300 test strips
Sample Throughput 210 samples per hour
Sample Identification Auto-ID barcode
Sample Rack 10-tube rack
Specimen Volume Min volume: 2.0 mL
aspiration volume: approx 1.0 mL
Sample Capacity Up to 210 samples with continuous feed
Data Storage 10,000 results
Computer Interface Ethernet
Operating Environment Temperature: 64° - 82°F (18° - 28°C)
Humidity: 20 - 80%; non-condensing
Power Supply 100 to 240 VAC; 50 to 60 Hz
Dimensions 22H x 21W x 24D in. (56H x 53.5W x 61D cm)
Weight 100 lbs. (45.4 kg)
footer